Medical Power Solution

IMA

12V 1000W / IMA-X1000-12

Highlights & Features

  • 500,000 hour MTBF
  • Safety rated for Medical, Industrial, IT and other applications
  • Wide operating input voltage range: 80 Vac to 275 Vac or 120 Vdc to 300 Vdc
  • Wide adjustable output voltage range
  • Low profile 1U design
  • Variable speed fan with low acoustic noise level of less than 39 dB(A)
  • High efficiency: up to 94%
  • Redundant operation with active current sharing
  • High isolation: 2 × MOPP
  • PMBus™ compatible for control, programming and monitoring
  • 3 years warranty
Dimensions (L x W x D) : 
209.5 x 127 x 40 mm
8.25" x 5.00" x 1.57"
Unit Weight : 
1.6 kg (3.53 lb)
Nominal Output Voltage
12V
Output Current
84A
Output Voltage Adjustment Range
9.6-14.4V
Output Power
1000W
PARD
< 120 mV
Start-up Time
< 2.5 s (@ AC OFF to ON, max. load)
< 150 ms (@ REMOTE OFF to ON, max. load)
Hold-up Time
20 msec nominal
Dynamic Response (Overshoot & Undershoot O/P Voltage)
< 5%
Nominal Input Voltage
100-240Vac
Input Voltage Range
80-275Vac (120-300Vdc)
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
DC Input Voltage Range
120 Vdc to 300 Vdc
Input Current
15A @ 80Vac
9.5A @ 120Vdc
Efficiency at 100% load
91% @ 115Vac
93% @ 230Vac
Max Inrush Current
< 20A
Power Factor at 100% load
0.9 (typical)
Dimensions
209.5 x 127 x 40 mm
8.25" x 5.00" x 1.57"
Unit Weight
1.6 kg (3.53 lb)
Cooling System
2 fans with variable speed control
Operating Temperature
-20°C to +70°C
Storage Temperature
-40°C to +85°C
Power De-rating
When AC input voltage is < 90 Vac, the output power will be reduced by 20 W per 1 V.
Operating Humidity
< 95% (non-condensing)
Shock Test (Non-Operating)
IEC 60068-2-27 compliant, 50 g, 11 msec, 3 shocks for each direction
Vibration Test (Non-Operating)
IEC 60068-2-6 compliant, 2.09 Grms, 5 Hz to 500 Hz, 20 minutes per side (3 planes)
Overvoltage
YES, latch mode
15V to 17.5V
Overload / Overcurrent
YES, at 108% to 140% of nominal output current; auto recovery
Over Temperature
YES, auto recovery
Short Circuit
YES, auto recovery
MTBF
500,000 hours (Telecordia SR-332, Issue 3, 25 °C, 90% confidence level)
Electrical Safety
UL 60950-1/CSA 22.2 No 60950-1, Edition 2; 5000 m (16,400 ft) altitude, 120 V to 300 Vdc and 100 V to 240 ±10% Vac (UL File E191395)
Medical Safety
IEC 60601-1(2005), EN60601-1(2006)
ANSI/AAMI ES 60601-1(2005)
CAN/CSA C22.2 No. 60601-1 (2008);
3,000 m (9,800 ft) altitude, 100 V to 240 Vac ±10% (UL File E325662)
Galvanic Isolation
Input to Output (Reinforced): 4000 Vac; 2 x MOPP
Input to Case (Basic): 1500 Vac; 1 x MOPP
Output to Case (Basic): 1500 Vac; 1 x MOPP
Emission
EN 55011, EN 55022 and FCC, Class B
Harmonic Current Emissions
EN 61000-3-2, Class A

Documents

App Notes Uploaded on 05-Jan-2018 3.5 MB
Datasheet Uploaded on 05-Jan-2018 2.7 MB
.DXF Uploaded on 05-Jan-2018 589 KB
.IGS Uploaded on 20-Oct-2015 14.1 MB
Manual Uploaded on 04-Jan-2018 855 KB
.STP Uploaded on 04-Jan-2018 910 KB
Image Uploaded on 03-Nov-2016 238 KB

Certificates

CB Uploaded on 05-Jan-2018 284 KB
CCC Uploaded on 05-Jan-2018 1.8 MB
CE Uploaded on 13-Feb-2018 444 KB
CE & RoHS Uploaded on 09-Oct-2015 223 KB
UL Uploaded on 17-Jan-2018 68 KB

Related Products

IMA 24V 1000W

IMA-X1000-24

-
Add to Comparison

Output Voltage

24V

Output Current

42A

Input Voltage

80-275 Vac

Dimensions

209.5 x 127 x 40 mm
8.25" x 5.00" x 1.57"

IMA 48V 1000W

IMA-X1000-48

-
Add to Comparison

Output Voltage

48V

Output Current

21A

Input Voltage

80-275 Vac

Dimensions

209.5 x 127 x 40 mm
8.25" x 5.00" x 1.57"

IMA 12V 600W

IMA-X600-12

-
Add to Comparison

Output Voltage

12V

Output Current

50A

Input Voltage

80-275 Vac

Dimensions

203.1 x 101.6 x 40.6 mm
8.00" x 4.00" x 1.60"

IMA 24V 600W

IMA-X600-24

-
Add to Comparison

Output Voltage

24V

Output Current

25A

Input Voltage

80-275 Vac

Dimensions

203.1 x 101.6 x 40.6 mm
8.00" x 4.00" x 1.60"
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT