Medical Power Solution

Medical Open Frame

12V 65W / MDS-065APS12 B

Highlights & Features

  • Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Low touch current (<0.1mA Normal & ,0.3mA single fault)
  • Over-Voltage/Load/Temperature & Short Circuit protections
  • 2.5 Million Hours MTBF
  • 2 x MOPP (means of patient protection)
  • 3 years warranty
Dimensions (L x W x D) : 
101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"
Unit Weight : 
0.13 kg (0.29 lb)
Nominal Output Voltage
12V
Output Current
5.42A
Output Power
65W
Line Regulation
± 0.5%
Load Regulation
± 5%
Start-up Time
3000ms @ 115Vac
Hold-up Time
15ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 3% @ 50-100% load
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
1.5A @ 100Vac,
0.75A @ 240Vac
Efficiency at 100% load
87.5% (typ.)
Max Inrush Current
30A @ 115Vac,
60A @ 230Vac (typ.)
Leakage Current
0.1mA @ 264Vac Normal Condition
, 0.3mA @ 264Vac Single Fault Condition
Dimensions
101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"
Unit Weight
0.13 kg (0.29 lb)
Operating Temperature
-10°C to +70°C
Storage Temperature
-40°C to +85°C
Power De-rating
-10°C to +50°C 100% load
50°C to 70°C de-rate power by 2.5% / °C
Operating Humidity
10-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minutes for each three axis
Overvoltage
150%, Latch Mode
Overload / Overcurrent
150%~210% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against Shock
Class I with PE* connection
MTBF
2.5 Million Hours based on Telecordia SR-332
CE
MDD Directive 93/42/EEC
Galvanic Isolation
Input to Output: 4000 Vac
Input to Ground: 1500 Vac
Output to Ground: 500 Vac
Emission
EN55011/22/FCC conducted / radiated class B
Voltage Fluctuation and Flicker
EN61000-3-3

Documents

Datasheet Uploaded on 18-Oct-2017 440 KB
.DXF Uploaded on 13-Jul-2015 255 KB
Mechanical Drawing Uploaded on 13-Jul-2015 25 KB
.STP Uploaded on 13-Jul-2015 741 KB
Image Uploaded on 03-Nov-2016 59 KB

Certificates

CB Uploaded on 21-Oct-2015 194 KB
CE Uploaded on 15-Aug-2016 116 KB
CQC Uploaded on 24-Apr-2018 246 KB
EAC Uploaded on 28-Jul-2016 1.2 MB
REACH Uploaded on 18-Dec-2018 441 KB
RoHS Uploaded on 18-Dec-2018 275 KB
UL Uploaded on 30-Mar-2018 421 KB

Related Products

Medical Open Frame 15V 63W

MDS-065APS15 B

-
Add to Comparison

Output Voltage

15V

Output Current

4.2A

Input Voltage

90-264 Vac

Dimensions

101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"

Medical Open Frame 18V 65W

MDS-065APS18 B

-
Add to Comparison

Output Voltage

18V

Output Current

3.61A

Input Voltage

90-264 Vac

Dimensions

101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"

Medical Open Frame 24V 65W

MDS-065APS24 B

-
Add to Comparison

Output Voltage

24V

Output Current

2.71A

Input Voltage

90-264 Vac

Dimensions

101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"

Medical Open Frame 12V 100W

MDS-100APS12 B

-
Add to Comparison

Output Voltage

12V

Output Current

8.33A

Input Voltage

90-264 Vac

Dimensions

127.0 x 76.2 x 31 mm
5.00" x 3.00" x 1.22"
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT