Medical Power Solution

MDS / MEA Adapter

24V 250W / MEA-250A24C

Highlights & Features

  • Meet Efficiency Level VI
  • Safety Approvals to IEC 60601-1 3.1 Ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 3th and 4th Ed. Requirements
  • IP22 ingress protection rating
  • 2 × MOPP isolation
  • Suited for Type BF Medical Products        
Dimensions (L x W x D) : 
200 x 100 x 43 mm
7.87" x 3.94" x 1.69"
Unit Weight : 
1.05kg (2.31 lb)
Nominal Output Voltage
24V
Output Current
10.42A
Output Power
250W
Line Regulation
± 0.5%
Load Regulation
± 4.5%
Start-up Time
3000ms @ 115Vac
Hold-up Time
10ms @ 100Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 10% @ 0-50% & 50-100% load; Slew rate 0.25A/μs (@ 100Hz, 50% Duty Cycle)
Case Chassis / Cover
PC
Dimensions
200 x 100 x 43 mm
7.87" x 3.94" x 1.69"
Unit Weight
1.05kg (2.31 lb)
Cooling System
NA
Operating Temperature
5°C to +60°C
Storage Temperature
-40°C to +70°C
Power De-rating
> 40°C de-rate power by 2.5% / °C
Operating Humidity
5-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minute for each three axis
Operating Altitude
5,000 meters (16,400 feet)
Overvoltage
(max.)
120%, Latch Mode
Overload / Overcurrent
(max.)
150% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode, (Non-Latching, Auto-Recovery)
Degree of Protection
IP22
Protection Against Shock
Class I
MTBF
(typ.)
500K Hours based on Telecordia SR-332 (at 115Vac, Max. load and 25°C Ambient)
Expected Cap Life Time
5 Years (115Vac & 230Vac, 100% load @ 25°C)
Electrical Safety
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2012, J60950-1(H29)
CE
MDD Directive 93/42/EEC
Medical Safety
IEC 60601-1 3.1 edition CB report
EN 60601-1
CAN/CSA-C22.2 NO. 60601-1:14,
ANSI/AAMI ES60601-1
UL60601-1 (File No. E356265)
Galvanic Isolation
Input to Output: 4000Vac
Input to PE: 1500Vac
Output to PE: 500Vac
Material and Parts
RoHS Directive 2011/65/EU Compliant
Emission
EN55011/EN55032, FCC Title 47: Class B
Immunity
IEC61000-4-2, IEC61000-4-3, IEC61000-4-4,IEC61000-4-5, IEC61000-4-6, IEC61000-4-8, IEC61000-4-11
Harmonic Current Emissions
IEC61000-3-2
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 06-Dec-2018 600 KB
.DXF Uploaded on 09-Feb-2018 154 KB
Mechanical Drawing Uploaded on 09-Feb-2018 135 KB
.STP Uploaded on 09-Feb-2018 415 KB
Image Uploaded on 08-Feb-2018 188 KB

Certificates

CB Uploaded on 09-Feb-2018 584 KB
CCC Uploaded on 09-Feb-2018 1.0 MB
CE Uploaded on 09-Feb-2018 156 KB
PSE Uploaded on 09-Feb-2018 2.4 MB
REACH Uploaded on 08-May-2018 868 KB
RoHS Uploaded on 31-Jul-2017 708 KB
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT