Configurable Power Supply

MEG

V 2100W / MEG-2K1A6

Highlights & Features

  • Up to 2100W in 5”x10”x1.59” Package
  • Up to 26.4W/inchPower Density
  • Full Power up to 50°C Ambient
  • Up to 500kHrs MTBF
  • 2xMOPP Isolation for Medical Application
  • Output selectable from 2V to 60V
  • Optional CV/CC Version on demand
  • Current sharing
  • Class B Conducted and Radiated EMI
  • IEC60601-1-2 4th edition immunity compliance
  • Normal and Reversed Option for Global Remote On/Off
  • Analog and Digital Voltage Trimming
  • PMBus Ver 1.3 Supported
  • Intelligent Fan Speed Control
  • Optional RS485/RS232/USB Communication Adapters
  • PC GUI for easy parameter setting and monitoring
Dimensions (L x W x D) : 
254 x 127 x 40.5 mm
10.00" x 5.00" x 1.59"
Unit Weight : 
2 kg (4.41 lb)
Output Power
2100W
Line Regulation
±0.5%
Load Regulation
±1%
Start-up Time
3000ms @ 115Vac
Hold-up Time
12ms @ 115Vac, with rated load
Dynamic Response (Overshoot & Undershoot O/P Voltage)
±5% @ with 50-100% load change
Rise Time
100ms
Nominal Input Voltage
100-240Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
15
Efficiency at 100% load
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz
Max Inrush Current
40A @ 230Vac, cold start
Power Factor at 100% load
>0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
see datasheet
Dimensions
254 x 127 x 40.5 mm
10.00" x 5.00" x 1.59"
Unit Weight
2 kg (4.41 lb)
Operating Temperature
Absolute Maximum/Minimum Rating.
-20°C to +70°C. Linearly de-rate from 100% load at 50°C, to 50% load at 70°C (2.5%/ degree centigrade)
Storage Temperature
-40°C to +85°C
Operating Humidity
5-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2Grms, 20 minute for each three axis
Operating Altitude
0-5,000 meters (0-16,400 feet)
Overvoltage
see datasheet
Overload / Overcurrent
see datasheet
Over Temperature
Latch off
Short Circuit
Hiccup Mode (Non-Latching, Auto-Recovery)
MTBF
>500kHrs based on Telecordia SR-332
Electrical Safety
IEC 60950-1 CB report
UL 60950-1+CAN/CSA 60950-1
CE
MDD Directive 93/42/EEC
Medical Safety
IEC 60601-1 3rd +A1 CB report
UL 60601-1+CAN/CSA 60601-1: (Ed.3.2005)
Galvanic Isolation
Input to/Output (2XMOPP): 4000 Vac
Input to/Ground (1XMOPP): 1500Vac
Output to/Ground: 500Vac (Type B application rated)
Material and Parts
RoHS Directive 2011/65/EU Compliant
Emission
EN55011/EN55032, FCC Title 47:Class B
Immunity
see datasheet
Harmonic Current Emissions
IEC61000-3-2: Meet Class A limit
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 06-Mar-2019 2.2 MB
.DXF Uploaded on 27-Aug-2018 916 KB
.STP Uploaded on 24-Aug-2018 13.8 MB
Manual Uploaded on 06-Mar-2019 4.1 MB
Image Uploaded on 21-Aug-2018 4.3 MB

Certificates

CB Uploaded on 22-Aug-2018 658 KB
CE Uploaded on 22-Aug-2018 35 KB
CSA Uploaded on 22-Aug-2018 77 KB
UPDATED REACH Uploaded on 20-Mar-2019 107 KB
RoHS Uploaded on 18-Dec-2018 275 KB
UL Uploaded on 22-Aug-2018 631 KB