Medical Power Solution

MDS / MEA Adapter

6V 5W / MDS-005AAS06 C

Highlights & Features

  • Meet Efficiency Level V
  • Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • IP22 Ingress Protection Rating
  • 1 Million Hours MTBF
Dimensions (L x W x D) : 
56.5 x 39.5 x 28 mm
2.22" x 1.56" x 1.10"
Unit Weight : 
0.1kg (0.22 lb)
Nominal Output Voltage
6V
Output Current
0.83A
Output Power
5W
Line Regulation
±1%
Load Regulation
±8%
PARD
111.2mV pk-pk @ Full load
Start-up Time
3000ms @ 115Vac
Hold-up Time
5ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
±9% @ 50-100% load
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
0.2A @ 115Vac ; 0.1 A @ 230Vac
Efficiency at 100% load
73% typ.
Leakage Current
0.1mA @ 264Vac NC, 0.3mA @ 264Vac SFC
Dimensions
56.5 x 39.5 x 28 mm
2.22" x 1.56" x 1.10"
Unit Weight
0.1kg (0.22 lb)
Case Chassis / Cover
PC / PC
Input Terminal
Wall mount - EU Type
Output Terminal
Barrel type
Operating Temperature
0°C to +40°C
Storage Temperature
-40°C to +85°C
Power De-rating
Linearly derate, at rate of 0.25 watts per degree C, from 5W at 40 degree C to 2.5W at 50 degree C
Operating Humidity
5-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minute for each three axis
Operating Altitude
3,000 meters (9842.52 feet)
Overvoltage
150% of rated output voltage (Non-latch Mode)
Overload / Overcurrent
130% of rated load current
(Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode
(Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode
(Non-Latching, Auto-Recovery)
Degree of Protection
IP22
Protection Against Shock
Class II
MTBF
1 Million hours based on Telecordia SR-332
Electrical Safety
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Medical Safety
IEC60601-1 3rd and 3.1rd editions CB reports
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012
Galvanic Isolation
Input to Output: 4000Vac
Material and Parts
RoHS Directive 2011/65/EU Compliant
Emission
EN55011/EN55032, FCC Title 47:Class B
Harmonic Current Emissions
IEC61000-3-2
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 28-Jun-2019 969 KB
Mechanical Drawing Uploaded on 21-Jul-2015 55 KB
Image Uploaded on 03-Nov-2016 93 KB

Certificates

CB Uploaded on 21-Jul-2015 564 KB
CCC Uploaded on 13-Jul-2015 681 KB
CE Uploaded on 11-Jul-2017 118 KB
REACH Uploaded on 20-Mar-2019 107 KB
RoHS Uploaded on 18-Dec-2018 275 KB
UL Uploaded on 21-Jul-2015 955 KB
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT