Medical Power Solution

MDS / MEA Adapter

19V 60W / MDS-060AAS19 B

Highlights & Features

  • Meet Efficiency Level VI
  • Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Low touch current (<0.1mA Normal & <0.3mA single fault)
  • Over-Voltage/Load/Temperature & Short Circuit protections
  • 1 Million Hours MTBF
  • 2 × MOPP isolation
  • 3 years warranty
Dimensions (L x W x D) : 
135 x 62 x 34.1 mm
5.31" x 2.44" x 1.34"
Unit Weight : 
0.36 kg (0.79 lb)
Nominal Output Voltage
19V
Output Current
3.15A
Output Power
60W
Line Regulation
± 0.5%
Load Regulation
± 4.5%
Start-up Time
3000ms @ 115Vac
Hold-up Time
10ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 5% @ 50-100% load
Ripple & Noise
61mV pk-pk @ Full load
Nominal Input Voltage
100-250Vac
Input Voltage Range
90-275Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
1.5A @ 115Vac,
0.75A @ 230Vac
Efficiency at 100% load
88.3% (typ.)
Max Inrush Current
50A @ 115Vac,
100A @ 230Vac (typ.)
Leakage Current
0.1mA @ 275Vac NC,
0.3mA @ 275Vac SFC
Dimensions
135 x 62 x 34.1 mm
5.31" x 2.44" x 1.34"
Unit Weight
0.36 kg (0.79 lb)
Case Chassis / Cover
PC
Operating Temperature
0°C to +40°C
Storage Temperature
-40°C to +85°C
Operating Humidity
10-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minute for each three axis
Operating Altitude
3,000 meters (9,840 feet)
Overvoltage
150%, Latch Mode
Overload / Overcurrent
180% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode, (Non-Latching, Auto-Recovery)
Degree of Protection
IP40
Protection Against Shock
Class I
MTBF
1 million Hrs based on Telecordia SR-332
Electrical Safety
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Medical Safety
IEC60601-1 3rd and 3.1rd edition CB report
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012
Galvanic Isolation
Input to Output: 4000Vac
Input to Ground: 1500Vac
Output to Ground: 500Vac
Material and Parts
RoHS Directive 2011/65/EU Compliant
Emission
EN55011, EN55032, FCC Title 47: Class B
Harmonic Current Emissions
IEC61000-3-2, Meet Class A limit
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 28-Jun-2019 1.1 MB
.DXF Uploaded on 14-Jun-2017 121 KB
Mechanical Drawing Uploaded on 13-Jul-2015 21 KB
.STP Uploaded on 09-Jun-2017 296 KB
Image Uploaded on 03-Nov-2016 49 KB