Medical Power Solution

MDS / MEA Adapter

12V 60W / MDS-060BAS12 A

Highlights & Features

  • Meet Efficiency Level VI
  • Safety Approvals to IEC 60601-1 3rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Low touch current (<0.1mA Normal & ,0.3mA single fault)
  • Over-Voltage/Load/Temperature & Short Circuit protections
  • 1 Million Hours MTBF
  • 2 x MOPP (means of patient protection)
  • 3 years warranty
Dimensions (L x W x D) : 
135 x 62 x 34.1 mm
5.31" x 2.44" x 1.34"
Unit Weight : 
0.34kg (0.75 lb)
Nominal Output Voltage
12V
Output Current
5A
Output Power
60W
Line Regulation
±0.5%
Load Regulation
±5%
PARD
84.8mV pk-pk @ Full load
Start-up Time
3000ms @ 115Vac
Hold-up Time
10ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 5% @ 50-100% load
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
1.5A@115Vac, 0.75A@ 230Vac
Efficiency at 100% load
88.3%
Max Inrush Current
50A@115Vac, 100A @ 230Vac
Dimensions
135 x 62 x 34.1 mm
5.31" x 2.44" x 1.34"
Unit Weight
0.34kg (0.75 lb)
Case Chassis / Cover
PC/PC
Operating Temperature
0°C to +40°C
Storage Temperature
-40°C to +85°C
Power De-rating
Linearly derate, at rate of 1.5 watts per degree C, from 60W at 40°C to 45W at 50°C
Operating Humidity
5 - 95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minute for each three axis
Operating Altitude
5,000 meters (16,400 feet)
Overvoltage
150%, Latch Mode
Overload / Overcurrent
180% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Degree of Protection
IP21
Protection Against Shock
Class II
MTBF
1 million Hours based on Telecordia SR-332
Electrical Safety
IEC60601-1: (Ed.3,2005),
EN60601-1:2006, CAN/CSA-C22.2 No. 60601-1:08,
ANSI/AAMI ES60601-1: (Ed.3,2005)
CE
MDD Directive 93/42/EEC
Galvanic Isolation
Input to Output: 4000 Vac
Emission
EN55011, EN55022, FCC Title 47: Class B

Documents

Datasheet Uploaded on 28-Jun-2019 763 KB
.DXF Uploaded on 14-Jun-2017 124 KB
Mechanical Drawing Uploaded on 21-Jul-2015 20 KB
.STP Uploaded on 14-Jun-2017 349 KB
Image Uploaded on 03-Nov-2016 69 KB

Certificates

CB Uploaded on 28-May-2019 571 KB
CCC Uploaded on 14-Dec-2018 1.1 MB
CE Uploaded on 15-Aug-2016 196 KB
REACH Uploaded on 20-Mar-2019 107 KB
RoHS Uploaded on 18-Dec-2018 275 KB
UL Uploaded on 21-Jul-2015 543 KB