Medical Power Solution

Medical Enclosed

12V 300W / MDS-300ADB12 AA

Highlights & Features

  • Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Up to 190 Watt Convection in 3.5" x 5.5" x 1.75" Package
  • Up to 800 K Hrs MTBF
  • Suited for Type BF Medical Products
  • 2 x MOPP isolation
Dimensions (L x W x D) : 
140 x 88.9 x 45 mm
5.5" x 3.5" x 1.75"
Unit Weight : 
0.61kg (1.34 lb)
Nominal Output Voltage
12V
Output Current
25A
Output Power
300W
Line Regulation
±0.5%
Load Regulation
±1%
Start-up Time
2000ms @ 115Vac
Hold-up Time
10ms @ 100Vac and 115Vac, with 300W load
20ms @ 100Vac and 115Vac, with 190W load
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 5% @ with 50-100% load change
Rise Time
100ms
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
4A
Efficiency at 100% load
91.3% @ 115Vac/60Hz
92.6% @ 230Vac/50Hz
Power Factor at 100% load
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
29.5uA @100Vac/60Hz(Typ)
81.5uA @264Vac/60Hz (Typ)
Dimensions
140 x 88.9 x 45 mm
5.5" x 3.5" x 1.75"
Unit Weight
0.61kg (1.34 lb)
Case Chassis / Cover
Enclosed cover
Operating Temperature
-10°C to +70°C
Storage Temperature
-40°C to +85°C
Power De-rating
Linearly derate from 100% load at 50°C, to 50% load at 70°C (2.5%/ degree centigrade)
Operating Humidity
5-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration (Operating)
5-500Hz, 2Grms, 15 minute for each three axis
Overvoltage
Main output 125% of rated voltage, Latch Mode
Standby 125% of rated voltage, Latch Mode
Overload / Overcurrent
Main output 160% of rated current
Standby 1.06A max
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode for Main output
Short Circuit
Hiccup Mode for Main and Standby output,
(Non-Latching, Auto-Recovery)
MTBF
800 Khrs based on Telecordia SR-332
CE
MDD Directive 93/42/EEC
Medical Safety
IEC60601-1 2nd and 3rd edition CB report TUV EN60601-1:2006 UL60601-1+CAN/CSA 60601-1: (Ed.3.2005)
Galvanic Isolation
Input to/Output (2XMOPP): 4000 Vac
Input to/Ground(1XMOPP): 1500 Vac
Output to/Ground(1XMOPP): 1500 Vac (Type BF application rated)
Emission
EN55011/EN55022,FCC Title 47:Class B
Harmonic Current Emissions
IEC61000-3-2 Meet Class D limit
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 25-Sep-2017 908 KB
Mechanical Drawing Uploaded on 27-Apr-2016 145 KB
.STP Uploaded on 12-Jul-2018 3.6 MB
Image Uploaded on 03-Nov-2016 914 KB

Certificates

CB Uploaded on 12-Jul-2016 174 KB
CCC Uploaded on 12-Jul-2016 14.1 MB
CE Uploaded on 16-Aug-2016 137 KB
UPDATED REACH Uploaded on 20-Mar-2019 107 KB
RoHS Uploaded on 18-Dec-2018 275 KB
TUV Uploaded on 12-Jul-2016 12.5 MB
UL Uploaded on 12-Jul-2016 1.7 MB
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT