Medical Power Solution

Medical Open Frame

18V 40W / MDS-040APS18 B

Highlights & Features

  • Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
  • Compliant with IEC 60601-1-2 4th Ed. Requirements
  • Low touch current (<0.1mA Normal & <0.3mA single fault)
  • Over-Voltage/Load/Temperature & Short Circuit protections
  • 3 Million Hours MTBF
  • 2 x MOPP (means of patient protection)
  • 3 years warranty
Dimensions (L x W x D) : 
76.2 x 50.8 x 23 mm
3.00" x 2.00" x 0.91"
Unit Weight : 
0.08 kg (0.18 lb)
Nominal Output Voltage
18V
Output Current
2.22A
Output Power
40W
Line Regulation
± 0.5%
Load Regulation
± 1%
Start-up Time
3000ms @ 115Vac
Hold-up Time
15ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 3% @ 50-100% load
Ripple & Noise
14.8mV pk-pk @ Full load
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current
0.85A @ 115Vac, 0.45A @ 230Vac
Efficiency at 100% load
86.9% (typ.)
Max Inrush Current
30A @ 115Vac, 60A @ 230Vac (typ.)
Leakage Current
0.1mA @ 264Vac Normal Condition
0.3mA @ 264Vac Single Fault Condition
Dimensions
76.2 x 50.8 x 23 mm
3.00" x 2.00" x 0.91"
Unit Weight
0.08 kg (0.18 lb)
Input Terminal
JST 2P Connector: B2P3-VH(LF)(SN)
Output Terminal
JST 4P Connector: B4P-VH(LF)(SN)
Operating Temperature
-10°C to +70°C
Storage Temperature
-40°C to +85°C
Power De-rating
-10°C to +50°C 100% load
50°C to 70°C de-rate power by 2.5% / °C
Operating Humidity
10-95% RH (Non-Condensing)
Shock Test (Non-Operating)
50G 11ms, 3 shocks for each direction
Vibration Test (Non-Operating)
5-500Hz, 2.09Grms, 20 minutes for each three axis
Operating Altitude
3,000 meters (9842.52 feet)
Overvoltage
150%, Latch Mode
Overload / Overcurrent
150%~210% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature
Hiccup Mode, (Non-Latching, Auto-Recovery)
Short Circuit
Hiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against Shock
Class I with PE* connection
MTBF
3 Million Hours based on Telecordia SR-332
Electrical Safety
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Medical Safety
IEC60601-1 3rd and 3.1rd edition CB report
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012
Galvanic Isolation
Input to Output: 4000 Vac
Input to Ground: 1500 Vac
Output to Ground: 500 Vac
Material and Parts
RoHS Directive 2011/65/EU Compliant
Emission
EN55011/EN55032, FCC Title 47: Class B
Harmonic Current Emissions
IEC61000-3-2
Voltage Fluctuation and Flicker
IEC61000-3-3

Documents

Datasheet Uploaded on 04-Mar-2019 605 KB
.DXF Uploaded on 13-Jul-2015 154 KB
Mechanical Drawing Uploaded on 13-Jul-2015 14 KB
.STP Uploaded on 13-Jul-2015 438 KB
Image Uploaded on 03-Nov-2016 54 KB

Certificates

CB Uploaded on 05-Mar-2019 1.8 MB
CCC Uploaded on 24-Oct-2017 3.1 MB
CE Uploaded on 15-Aug-2016 118 KB
UPDATED REACH Uploaded on 20-Mar-2019 107 KB
RoHS Uploaded on 18-Dec-2018 275 KB
UL Uploaded on 30-Mar-2018 392 KB

Related Products

Medical Open Frame 12V 40W

MDS-040APS12 B

-
Add to Comparison

Output Voltage

12V

Output Current

3.33A

Input Voltage

0-0 Vac

Dimensions

76.2 x 50.8 x 23 mm
3.00" x 2.00" x 0.91"

Medical Open Frame 15V 40W

MDS-040APS15 B

-
Add to Comparison

Output Voltage

15V

Output Current

2.67A

Input Voltage

0-0 Vac

Dimensions

76.2 x 50.8 x 23 mm
3.00" x 2.00" x 0.91"

Medical Open Frame 24V 40W

MDS-040APS24 B

-
Add to Comparison

Output Voltage

24V

Output Current

1.67A

Input Voltage

0-0 Vac

Dimensions

76.2 x 50.8 x 22.9 mm
3.00" x 2.00" x 0.90"

Medical Open Frame 15V 63W

MDS-065APS15 B

-
Add to Comparison

Output Voltage

15V

Output Current

4.2A

Input Voltage

0-0 Vac

Dimensions

101.6 x 50.8 x 30 mm
4.00" x 2.00" x 1.18"
We use cookies to provide the best user experience for those who visit our website. By using this website you agree to the placement of cookies and our Privacy Policy . ACCEPT